s Septech

Optizen Nanoq Lite

OPTIZEN NanoQ Lite is a small micro sample analyzer with a simple yet sophisticated design and easy and convenient user interface (UI).

It applies the Slope Algorithm, which is a light path optimization technique, to enable measurement of a wide range of absorbance.


Main Charateristics

  • It can measure protein and conduct cell counting.
  • The measurement result can be saved or downloaded to a USB flash drive or smartphone.
  • A user can easily operate the system using the intuitive LCD screen and touch functions.
  • It does not need a separate computer since the main unit performs the control.
  • The measurement data can be saved in the equipment to improve the spatial utilization and economic efficiency.
  • The lightweight and small spectrophotometer features the dimensions of 148 x 179 x 110 mm and weight of 1.4 kg.
  • The built-in touch LCD panel helps users operate it easily and conveniently.

Product details

  • Using three LED (260 nm, 280 nm, and 600 nm) and a single silicon photodiode, it lowered the product cost drastically while maintaining the needed functions and performance level.
  • It can perform the baseline correction function using an additional LED (360 nm). There is no need for a user manual since it provides a simple and intuitive user environment. The designated factor is set as the default value for each mode, and the input mode is provided to allow the users to set a factor value which is different from the default.
  • Users can specify the factor appropriate for the user preference, the lab environment, and the type of protein.

Product Information

  • It is a small spectrophotometer featuring compact size and main unit control.
  • There is no need for measurement standby since it does not require a lamp warming-up.
  • A user can conduct the measurement quickly since there is no need to set the pathlength according to the concentration.
  • A user can run the main functions quickly with a button.

Specifications

Instrument Type Microvolume Spectrophotometer
Minimum Sample Volume 1 μL
Sample Number 1
Pathlength 0.03 ~ 0.5 mm (Auto ranging)
Spectral Resolution ≤ 8.0 nm
Absorbance Precision 0.002 Abs (0.5 mm pathlength)
Absorbance Range 0 – 40 Abs
Detection Limit About 2 ng/μL (dsDNA)
Maximum Concentration 2,000 ng/µL (dsDNA), 60 mg/ml (BSA), 28.8 mg/ml (IgG)
Photometric Range(10 mm Equivalent) Cuvette: 0~2 A, Accuracy - 3% (at 1A/mm at 280 nm)
Photometric Accuracy 3% (at 1A/mm at 280 nm)
Light Source LEDs
Measurement Wavelenght 260, 280 nm / 600 nm(Cuvette)
Baseline Correction Wavelenght 360 nm
Detector Silicon photodiode
Measurement Time ‹10 seconds
Footprint(W × D) 145 x 190 mm
Weight 1.4 kg
Operating Voltage 12 V (DC)
Power Consumption Operating - 4.7~5.2 W, Standby - 3.3 W
Display 4.3-inch, Touch-screen glove compatible
Display resolution 480 x 272 pixels (Color display)
Internal Storage 4 GB
Conectivity USB-A, USB-B, RS232C, Bluetooth
Accessories Thermal Printer(Optional)